7 Alarms and messages
7-12
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
A
larm
ID
Prio. Device
action
Cause
Remedy
Low etVA
[09] High --
The etVA
concentration has
fallen below the
lower alarm limit,
while the VA
application is active
(
Start
button in
active status).
Check patient condition.
Check alarm settings.
Check, if VA application is
active.
Check MIRUS Reflector
and interface.
Check MACset.
Low Vti
[04] Low
--
Measured Vt
i
is
lower than 200 mL.
Check patient condition.
Check ventilator settings
(Vti has to be min 200 mL).
7.3.2 Technical alarms
Alarm
ID
Prio. Device
action
Cause
Remedy
Device
inclined
[21] Low
Pauses VA
delivery.
MIRUS Controller is
tilted.
Reposition MIRUS
Controller on level surface.
Device
inclined!
[22] High Pauses VA
delivery.
MIRUS Controller is
tilted.
Reposition MIRUS
Controller on level surface.
Dosage failed!
Start VA?
[16]
[17]
[18]
[19]
[50]
High Stops VA
delivery.
Calculation problem
was detected.
Restart VA delivery by
pressing Play and
confirmation or turn off
MIRUS Controller.
Failure!
Turn off.
[25]
[26]
[27]
[28]
[29]
[30]
[31]
High Stops VA
delivery.
A technical failure
was detected.
Turn off MIRUS Controller
and restart.
If alarm persists, call
service.
Fill port flap
still open
[37] Low
--
Fill port flap is open
for more than
5 min.
Close fill port flap.
Fill port flap
still open!
[38] High --
Fill port flap is open
for more than
7 min.
Close fill port flap.
Heating system
failed
[48] High --
No activity of the
heating system
could be detected.
Turn off MIRUS Controller.
Call service.
Interface
disconnected
[39] High Pauses VA
delivery.
MIRUS Reflector
plug is removed
from the receptacle.
Reconnect the MIRUS
Reflector plug.
Mains supply
lost
[33] Low
--
MIRUS Controller
lost mains power
supply.
Connect the MIRUS
Controller to the mains
power supply.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...