8 Specifications
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
8-11
8.8.2 Specification of Wash-in speed parameters
Volatile Anaesthetic
Specification Wash-in speed
parameters
Note
Isoflurane
Cheetah = Fast.
2* set MAC.
Rabbit = Medium.
1.5* set MAC.
Turtle = Slow.
1.0* set MAC.
max. concentration: 5.0 Vol%.
Sevoflurane
Cheetah = Fast.
2* set MAC.
Rabbit = Medium.
1.5* set MAC.
Turtle = Slow.
1.0* set MAC.
max. concentration: 8.0 Vol%.
Desflurane
Cheetah = Fast.
1.5* set MAC.
Rabbit = Medium.
1.25* set MAC.
Turtle = Slow.
1.0* set MAC.
max. concentration: 12.0 Vol%.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...