6 Cleaning and maintenance
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
6-5
Affected part
Period
Performer
Fan filter
Replace at least every 12 months.
User / Operator
Software
Manufacturer gives information as
soon as software update is available.
User / Operator
Comment:
For maintenance report template and for further information to the tests please contact the
manufacturer TIM GmbH via: [email protected]
Visual inspection
Inspect the MIRUS Controller for distortion, deterioration and cracks:
Check safety relevant labelling (SN/SI-Label) on the device for availability and
readability.
Check mechanical parts are not defective.
Check for damage and debris.
Check relevant accessories (original power supply cord).
Check accompanying product documentation, IFU and labelling is available and
matches version of product.
Electrical safety test
In this test the following parameter are tested according to IEC 62353:2014 (base is
IEC 60601) (class I, B):
Protective earth resistance
(PASS < 0.2
)
Earth leakage current
(PASS < 500 µA)
Insulation resistance
(PASS > 10 M
)
Take detachable power supply cord into measurement.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...