4 Preparation
4-10
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
7. Take the bottle with the adapter and
insert it vertically, with the adapter facing
the Fill port.
Press the bottle into the Fill port gently
until you feel a mechanical stop that
opens the spring loaded valves.
8. Watch the screen and wait until you read
the message “The maximum level is
reached”. Remove the bottle and close
the Fill port flap manually by pushing it
down until it locks in.
Comment:
Device cannot be overfilled.
If the content of the bottle is emptied
before the maximum level is reached,
simply continue with a new bottle of
anaesthetic agent until maximum level is
reached and close the Fill port flap.
9. The MIRUS Controller will now resume
start up procedure provided that
anaesthetic was refilled and Fill port flap
is closed properly.
IMPORTANT
Risk of equipment damage
To avoid the ingress of particles and
other substances, close the fill port flap
immediately after filling.
HW 2.n
SW 2.03.02
The maximum level is reached.
Please close the Fill port.
HW 2.n
SW 2.03.02
Thank you.
Power up test will now proceed automatically.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...