7 Alarms and messages
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
7-11
7.3 Alarm messages
The MIRUS Controller distinguishes between patient related alarms and technical alarms.
7.3.1 Patient alarms
A
larm
ID
Prio. Device
action
Cause
Remedy
Apnoea
[01] Low
Pauses VA
delivery.
Patient stopped
breathing.
Obstruction
Check patient condition.
Check interface for kinking.
Check ventilator settings.
High etCO
2
[12] High --
The upper alarm
limit for the etCO
2
concentration has
been exceeded.
Check patient condition.
Check alarm settings.
Check ventilator settings.
High etVA
[10] High --
The upper alarm
limit for the etVA
concentration was
exceeded while the
VA application is
active (
Start
button
in active status).
Check patient condition.
Check alarm settings.
Check MACset.
Low etCO
2
[11] Low
--
The lower alarm
limit for the etCO
2
concentration has
been exceeded.
Check patient condition.
Check alarm settings.
Check ventilator settings.
If an ECMO or ECCO2R
therapy is being
administered: check the
settings.
Low etCO
2
[13] High Pauses VA
delivery
The etCO
2
concentration has
fallen below the
threshold value of
12 mmHg.
Check patient condition.
Check ventilator settings.
If an ECMO or ECCO2R
therapy is being
administered: check the
settings.
Check the MIRUS Reflector
for leakage.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...