6 Cleaning and maintenance
6-8
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
Safety function test
In this test all safety relevant functions of the MIRUS Controller are tested by the device itself.
The following accessory is required (refer to Chapter 6.6):
MIRUS Reference MR
Perform the safety function test as follows:
a. Mount the MIRUS Reference MR and perform a Power up Test and a System test
according to Chapter 4.
b. Safety function test is PASS when device finishes testing with the on screen
message “Power up test and System test successfully completed on (date) at
(time). To continue please choose a menu button on the right".
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...