6 Cleaning and maintenance
6-4
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
6.2 Maintenance
General information
WARNING
Risk to persons from volatile anaesthetic in the MIRUS Controller
Volatile anaesthetic may leak out and contaminate the environment.
Empty out the MIRUS Controller before maintenance, as described
in Chapter 6.3.
WARNING
Infection risk
The MIRUS Controller may be contaminated by pathogens.
Clean the MIRUS Controller before maintenance, as described in
Chapter 6.1.
WARNING
Risk to persons and equipment damage due to incorrect
maintenance
Maintenance measures may only be conducted by maintenance
personnel in accordance with Chapter 3.2.
Schedule
Affected part
Period
Performer
MIRUS Controller
Every 12 months Safety Check
Visual Inspection
Electrical Safety Test
Safety Function test
Manufacturer or qualified
personnel
MIRUS Controller
Every 24 months Safety Check
Visual Inspection
Electrical Safety Test
Checking gas monitor
Checking pump
Safety Function test
Manufacturer or qualified
personnel
Emergency battery
(UPS battery)
When not using the MIRUS Controller,
leave MIRUS Controller connected to
mains supply or connect to mains
supply at least once a month for at
least 24h.
User / Operator
Emergency battery
(UPS battery)
Replace every 24 months.
Manufacturer or qualified
personnel
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...