8 Specifications
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
8-13
Danger due to incomplete data transfer
Data transmitted to the patient monitor, such as measured values or alarms, may be displayed
incompletely or incorrectly. The data is used as complement of the MIRUS Controller as
primary alarm source to establish a distributed alarm system with unconfirmed alarm
transmission according to IEC 60601-1-8. Only use the MIRUS Controller as the primary
source for alarms and viewing measured values.
Danger due to delayed data transfer
Signals from measured values and alarms of the MIRUS Controller can be delayed before
transfer to a patient monitor. These delays may vary depending on the conditions under which
the product is operating. For this reason, specified delay times are not available.
8.9.3 RJ22/RJ10 interface
The RJ22/RJ10 interface transfers a high priority alarm signal to a connected nurse call device.
The transfer takes place via a galvanic insulated changeover contact. Data transfer to the
MIRUS Controller is not possible.
Wiring diagram:
8.9.4 LAN interface (RJ45)
The LAN interface uses TCI-IP and is only used for service purpose by authorized personnel.
For safety reasons, this interface is deactivated during the application of volatile anaesthetic.
8.9.5 SD card interface
Software updates of the MIRUS Controller can be carried out via the SD card interface (micro
SD). For security reasons, a dedicated and coded SD card (SW update) is needed, which the
manufacturer (TIM) makes available as required.
Comment:
For further information about the software update refer to the Additional Information (AI)
Chapter 6.3.
Event log data of the MIRUS Controller can be downloaded via the SD card interface (micro
SD), e.g. for scientific purposes. For security reasons, a dedicated and coded SD card (Data
download) is needed, which is available from the manufacturer (TIM).
Pin 2
Pin 4
Pin 1
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...