7 Alarms and messages
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
7-5
7.2.2 Audible alarm escalation
Comment:
The alarm volume will escalate from 50% sound intensity up to the set volume (max. 100%)
until alarm is muted.
Comment:
The set value for alarm volume represents the maximum sound level.
First time of alarm appears
Second time of alarm appears
since 2 sec
Low etCO
2 [11]
since 12 sec
Low etCO
2 [11]
etSevo: 1.9
Vol%
etCO
2
: 25
mmHg
etSevo: 1.9
Vol%
etCO
2
: 25
mmHg