2 Safety
2-10
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
2.2 Fail-safe mode
Safe mode means:
•
screen is blank
•
no volatile anaesthetic agent is
delivered
•
alarm lights are flashing in red
•
confirm button is flashing in red
•
audible alarm sound
To silence the audible alarm, press the Confirm button for at least 4 seconds. MIRUS Controller
now turns off flashing red LEDs and stops alarm sound. To turn off the fail-safe mode
disconnect mains supply and press the Confirm button again for at least 4 seconds. The red
LEDs now turn off.
To restart MIRUS Controller after a fail-safe mode
•
wait for 5 seconds,
•
reconnect AC supply,
•
re-start by pressing the Confirm button (usual start).
2.3 Notification of serious incidents
The manufacturer and the responsible agency in the state in which the MIRUS Controller is
being operated must be notified of any serious incidents occurring in connection with the
MIRUS Controller.
Serious incidents include device-related death and severe injuries or illnesses.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...