6 Cleaning and maintenance
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
6-3
Cleaning the individual components
MIRUS Controller surface
a. Turn off the MIRUS Controller and ensure that the mains power cord is
disconnected.
b. Make sure the Fill port is closed.
c. Remove MIRUS Reflector and interface plug.
d. Make sure the receptacle token is placed in the MIRUS Reflector receptacle and
the receptacle is sealed.
e. Make sure the three connector tokens are placed in the MIRUS Controller
communication ports and the ports are sealed.
f. Carry out wipe disinfection.
g. Allow the device surface to dry fully.
MIRUS Controller Fill port
If the fill port is contaminated, clean it using a dry, clean, lint-free cloth.
IMPORTANT
Risk of equipment damage
Do not use cotton swabs for cleaning in order to prevent the ingress of small particles and lint.
MIRUS Reflector
To clean the outer surfaces of the MIRUS Reflector, follow the instructions supplied with the
cleaning product.
MIRUS Filter
The MIRUS Filter is a single use device and should not be soaked, rinsed, washed, sterilised
or treated with liquid disinfectants.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...