High priority alarms
MIRUS Controller, SW 2.03.02, Additional Information Master (en-GB)
BC-2
High priority alarms
Alarm
What MIRUS Controller does
What to check or to do
Device
inclined!
Pauses the application and closes
all valves until MIRUS Controller is
positioned on a level surface.
1. Reposition MIRUS Controller
on a level surface.
Dosage failed!
Start VA?
Stops the application.
1. Restart VA delivery by
activating Play and Confirm or
turn off MIRUS Controller.
Failure!
Turn off.
Stops the application.
Turn off tag becomes selectable.
A major technical failure was
recognised.
1. Turn off MIRUS Controller.
2. Wait 5 seconds and try to start
MIRUS Controller again.
3. If during self-test "Sorry…"
message appears the MIRUS
Controller is seriously
damaged. Call your service
organisation.
Fill port flap
still open!
Continues working.
1. Close Fill port flap.
Heating
system failed.
Continues working.
1. Turn off MIRUS Controller.
High etVA
Continues working.
1. Patient okay (CO
2
, SpO
2
)?
2. Set MAC okay?
3. Alarm settings okay?
High etCO
2
Continues working.
1. Ventilator settings okay?
2. Alarm settings okay?
3. Patient okay?
Interface
disconnected
Pauses the application.
1. Reconnect MIRUS Reflector
interface plug.
Low etCO
2
Pauses the application.
1. Ventilator settings okay?
2. Alarm settings okay?
3. Patient okay?
4. Leakage from the MIRUS
Reflector?
5. Are ECMO / ECCO2R settings
ok?
Low etVA
Continues working.
1. Delivery active (Start button!)?
2. Alarm settings correct?
3. MRUS Reflector interface
okay?
Occlusion of
Interface
Pauses the application.
1. MIRUS Reflector interface bent
or pinched?
2. Is fresh gas flow in the
anaesthesia device at least 1.5
x MV?
Reservoir
empty
Continues working.
1. Refill reservoir promptly.
UPS battery
low
Continues working.
1. Recharge battery.
Содержание Mirus MC DES
Страница 1: ...Additional Information MIRUSTM Controller Software version 2 03 02 Hardware version 2 n...
Страница 28: ...1 Introduction 1 22 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 48: ...3 General description 3 8 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 78: ...5 Operation 5 18 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 92: ...6 Cleaning and maintenance 6 14 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...
Страница 118: ...7 Alarms and messages 7 26 MIRUS Controller SW 2 03 02 Additional Information Master en GB Notes...