3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
167
Vial Option Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Depenetration Speed
This parameter allows you to select the required speed for
pulling the syringe needle out of the reservoir vial.
Diameter
Specific for Waste Vial.
Defines the diameter of the reservoir container.
Height
Specific for Waste Vial.
Defines the height of the reservoir container.
Penetration Speed
The syringe needle penetration speed is set by this value.
It is recommended to use the default value first.
Changing this parameter in one or the other direction might
lead to needle bending.
Vial Type
A list with all predefined Vial Types opens up for selection.
Special cases for
Waste Vial
:
•
None
— It is selectable, however if selected in
combination with a Wash Module is an error message
(Pending Message) propagated. Incompatible Type.
•
Waste Port
—
Used for the Fast Wash and Large Wash
Module.
•
Waste Vial
—
Used for Standard Wash Module.
•
Waste Inset
—
This type is used for Standard Wash
Module with inserted Waste Line Adapter.
Volume
Specific for
Wash
and
Waste Vial
.
Defines the volume of the reservoir container.
Z-Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) where the Head must expect an object.
• If the value of this item is
> 0
, indicates a relative
detection mode, and the sensor from the Head needle
guide is active.
• If the value of this item is set to
0
the sensor of the
needle guide is turned off and the syringe slider moves
to an absolute value, the Z-axis position as specified by
the object.
Table 78.
Vial Parameters (Sheet 2 of 2)
Item
Description