3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
125
Figure 85.
Menu List Item Signals and List Items Input and Output Signals
Press
Enter
to edit a parameter. See
Table 55.
Input Signal Parameters
Items
Description
Input Signal
Selects the required Input Signal. Typically used for the
Ready Signal (TTL-In).
Signal On
Indicates the signal has been received. A crossed-out box
means that a signal has been received.
Blocking Time
Used to program a time delay after receiving the signal
(usually the
Ready Signal
from an external device).
Source
Selects the required Input Signal for synchronization with an
external device.
Usually a
TTL-In
contact is used to receive the
Ready
Signal
from an external device.
Table 56.
Output Signal Parameters (Sheet 1 of 2)
Item
Description
Output Signal
Selects the required Output Signal
Signal On
Indicates if the signal has been received. A crossed-out box
means that a signal has been received.
Destination
Selects the required Output signal for the synchronization
with an external device.
Usually a switch out signal (SW Out) contact is used to send
the
Start Signal
to an external device.