Service Menu Item
Installation Item
208
TriPlus RSH User Guide
Thermo Scientific
3. Adjust the
Fiber Alignment
. Align the fiber tip with the needle tip as indicated in the
figure of the handheld controller, or the virtual handheld controller. Turn the fiber
downwards until it is clearly visible. Turn it back at the same level as the needle tip, then
save this value. See
Figure 148.
SPME Fiber Alignment
4. Adjust the
Standby Position
. Bring back the fiber by at least 2 mm into the needle
(protection sheet), so the fiber does not touch the septum when penetrating through.
The fiber could stick to the septum and be damaged or even lost. See
.
Figure 149.
SPME Fiber Standby Position
For the two versions an approximate value for the parameter
Fiber Alignment
and the
recommended value for the parameter
Fiber Standby Position
are given in
Figure 150.
Fiber Alignment and Fiber Standby Position
SPME Type
Fiber Alignment
Fiber Standby Position
SPME 1
10.8 mm
2.0 mm
SPME 2
3.4 mm
2.0 mm