Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
173
Wash Stations Option Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
for
Standard
and
Large Volume Wash Stations
, and
for
Fast Wash Station
.
Table 82.
Standard and Large Volume Wash Option Parameters
Item
Description
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly.
In case a teaching point is incorrect, return to the Menu
Setup
for readjustment of the teaching position. Please note
that this action can be done under the
Extended User Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again
(teaching).
New Wash Station
Creates of a new Wash Station.
Copy Wash Station
Copies of the present Wash Station.
Delete Wash Station
Deletes of the present. Wash Station.
Table 83.
Fast Wash Option Parameters
Item
Description
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again
(teaching).
Reset to Default
Resets the selected station parameter to default value.
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly.
In case a teaching point is incorrect, return to the Menu
Setup
for readjustment of the teaching position. Please note
that this action can be done under the
Extended User Level
but not under
User Level
.
New Wash Station
Creates of a new Wash Station.
Prime Wash Liner
Primes the tubing of the selected Wash Position.
Duration
Sets the time required for priming.