3
Setup Menu Item
Setup Menu Item Modules
118
TriPlus RSH User Guide
Thermo Scientific
Sample Drawers
Selecting the list item
Sample Drawers
opens up the following parameters. See
Figure 83.
Menu Item Temperature Controlled Drawer
and Menu List Item Stack
Press
Enter
to edit a parameter. See
Check Teaching
Checks the teaching position of the module.This task serves
as a quick check to determine if the X-, Y-, and Z-positions
are defined correctly. In case a teaching point is incorrect,
return to the Menu
Setup
for readjustment of the teaching
position. Please note that this action can be done under the
Extended User Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again,
(teaching).
Table 48.
Obstacle Option Parameters (Sheet 2 of 2)
Item
Description
Note
The Temperature Controlled Drawer can be adjusted through the BUS protocol.
For the power supply it is mandatory to use a second Power Module. Do not use the same
power supply as used for the TriPlus RSH.
Table 49.
Stack Module Parameters (Sheet 1 of 3)
Item
Description
Drawer #
This parameter is for information only.
The status of the drawer can be checked in
Drawer Status
.
Actual Temperature
This parameter is for information only.