5
Service Menu Item
Installation Item
204
TriPlus RSH User Guide
Thermo Scientific
The wizard
Teach Tool Station
calibrates the position of the slot(-s). The height of the tool,
installed in the slot, is calibrated against the TriPlus RSH Injection Unit base plate to define
the coupling distance.
To calibrate the position of the slots
1. Select the wizard item
Teach Tool Station
. See
.
Figure 142.
Wizard Item Teach Tool Station
The wizard automatically turns off the current to the TriPlus RSH to enable a manual
movement of the TriPlus RSH Injection Unit to the first Slot.
If you have performed
, a tool should still be installed.
If for any reason there is no installed tool, the TriPlus RSH will recognize the situation
and will go through an intermediate step to mount a tool.
WARNING
When an ATC Station is used in the configuration of the system, remove any
other tool from the slots before the wizard is started. If this is not observed, an error
message will appear.