1
Getting Familiar with Your TriPlus RSH
Temperature Controlled Drawer
74
TriPlus RSH User Guide
Thermo Scientific
Temperature Controlled Drawer Specifications
The specifications are:
•
Sample Capacity —
Accommodate two different sample trays.
Compatible VT70, VT54, and VT 15 (10 mL vials only).
•
Total vial height
— 47 mm (including cap)
•
Temperature Control
— From 4 °C – 40 °C in 0.1 °C increments
Peltier element for the cooling device
•
Temperature Cooling
— The specification for the Temperature Controlled Drawer
temperature cooling is defined as:
TΔ = ≥ 14.0 °C within 60 minutes.
TΔ = T Room temperature – T Temperature Controlled Drawer
•
Electrical Connectors
— Two BUS connectors for daisy chaining and a connector for
external power supply.
The Temperature Controlled Drawer dimensions, mass, operating and environmental
requirements are detailed in
.
Table 26.
Temperature Controlled Drawer - Dimensions and Mass
Length
Dimension in mm (in.)
Temperature Controlled Drawer
Width
197 (7.78)
Depth
425 (16.73)
Height
103 (4.06)
Power Supply
Width
135 (5.31)
Depth
215 (8.46)
Height
75 (2.95)
Mass
Dimension in kg (lbs)
Temperature Controlled Drawer
3.57 (7.87)
Power Supply
1.30 (2.87)
Associated cables (3 pieces)
0.50 (1.10)