3
Setup Menu Item
Setup Menu Item Modules
102
TriPlus RSH User Guide
Thermo Scientific
Select
Vortexer
, then press
Enter
to edit a parameter. See
Vortexer Options Pull-up Menu
Selecting the pull-up menu
Options
shows the following tasks. See
.
Table 33.
Vortexer Parameters
Item
Description
Actual Position
Reads out the position of the Agitator; read-only function.
Table 34.
Vortexer Pull-up Menu Options
Item
Description
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly. In case a teaching
point is off, return to the Menu
Setup
for readjustment of
the teaching position.
Please note that this action can be done under the
Extended
User Level
but not under
User Level
.
Teach Module
The positions of the X-, Y-, Z-axes can be defined again, for
teaching.
Enable Motor
Enables the motor of the vortexer.
Disable Motor
Disables the motor of the vortexer.
Check Vortexer
Starts the spinning motion of the Vortexer. It is used to test
the mixer function or the spinning speed for particular
application.
Settings are:
•
Option
— On; Off; Home
•
Speed
—
Select the required speed, range from 250 rpm
to 2500 rpm
Press
Run
to start the Vortexer
To stop the motion, press
Options again
, select
Check
Vortexer
and set the parameter
Option
to
Off
. Press
Run
.
Содержание TriPlus RSH