3
Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
149
Slot Offset X
Distance from Tray Holder Zero Point to Slot in X-axis
direction.
Used for Trays which are larger than the footprint of the
Slot, (Oversize).
This parameter is for example used for the Tray Type R60.
Value = -13.5 mm.
Slot Offset Y
Distance from Tray Holder Zero Point to Slot in Y-axis
direction.
Used for Trays which are larger than the footprint of the
Slot, (Oversize).
This parameter is for example used for the Tray Type R60.
Value = -13.5 mm.
Slot Offset Z
Distance from Tray Holder Zero Point to Slot in Z-axis
direction.
Used for Trays which are larger than the footprint of the
Slot, (Oversize).
This parameter is for example used for the Tray Type R60.
Value = -3.8 mm.
Slot Span
Number of the Tray installed on the Tray Holder.
Examples
: Tray 60: Slot Span 1. Tray 54: Slot Span 3.
Note
: A Tray Type with a
Slot Span >1
has to be positioned
in Slot 1. The Rack does
span
from Slot 1 to Slot 2 to Slot 3.
Total Column Length X
Defines the total length of the Column (X-axis). Distance
from center to center of cavity is measured. Unit: mm.
Total Row Length Y
Defines the total length of the Row (Y-axis). Distance from
center to center of cavity is measured. Unit: mm.
Uses Adapter
Select this parameter when an Adapter is used to compensate
for vial height. The adapter installed plus the vial height
must not exceed the total height of a 20 mL vial.
Example
: 10 mL vial plus corresponding adapter must be
equal to the height of a 20 mL vial.
Table 70.
Tray Type Parameters (Sheet 2 of 3)
Item
Description