1
Getting Familiar with Your TriPlus RSH
Barcode Reader
66
TriPlus RSH User Guide
Thermo Scientific
Approved Barcode Labels
The Barcode labels should be made of polyester and not paper. Polyester can withstand high
temperatures (Agitator 200 °C maximum) and the barcode lines will be printed clearly.
The minimum length of the barcode is 18 mm if used with a 20 mL vial. This dimension
refers to the actual barcode length and not the label itself. Adapt the label accordingly. Please
note that this is the minimum length; when possible use larger dimensions for reliable
processing of the barcodes.
Barcode Label Positioning
Place the barcode label on the vial such that the barcode bars are horizontally positioned.
•
Maximum tilt of the barcode label
: ±20°.
•
Label Width
: The barcode label width (length of the bars/spaces) has to be at least
18mm.
•
Minimal barcode density (minimal width of a bare or a space)
: 5 mil (0.005”) /
0.127mm.
The allowed area for the placing of the label is given in
.
Code 39 (3 of 9)
Alphanumeric code, consisting of
uppercase letters (A-Z), numeric digits
(0-9) and some special characters (-, .,
$, /, +, %, and space).
1
Unlimited
2 of 5 Interleaved
Numerical characters.
1
Unlimited
ISBT 128
Used for labeling of human blood.
1
Unlimited
Application specific barcodes
Table 20.
Barcode Symbologies (Sheet 2 of 2)
Barcode Type
Description
Min Length
Max
Length
Barcode Symbology