Getting Familiar with Your TriPlus RSH
Control Interface
34
TriPlus RSH User Guide
Thermo Scientific
•
S1
— Reset button to restore the factory default settings without knowing the IP
Address.
•
HANDHELD CONTROLLER
— A 20-pin port connector for the connection between
the TriPlus RSH and the Handheld Controller.
•
USB Host
— USB Type A connector for the connection of the USB stick only for
service purposes.
•
USB Device/Client
— USB Type connector for the connection to an USB device
or USB client.
•
Ethernet
— RJ45 Ethernet cable connects the TriPlus RSH directly to a computer,
and a LAN (Local Area Network) for the network connection of the TriPlus LAN
sampler.
•
BUS
— Two Mini Delta Ribbon 14-pin connectors for the connection to the CPU board
and the components installed on the sampler.
•
INTERFACE
— DSB 15-pin female connector for the connection between the sampler
and the GC.
•
FUSE
— Housing for the fuse T6.3; 250 VDC. See
.
•
POWER
— KYon KPJX-3S-S connector receives 36 V from the Power Module.
Blue status light, blinking:
The TriPlus RSH is in the process of software
Update
or
Backup
. (For the processes
Copy Backup
and
Restore
the Status
light remains
green
).
Blue status light, permanent:
TriPlus RSH is in a safe state (power reduced to
a minimum) and requires manually moving the Head to the
Teach Point
.
Note:
If an error is detected at the same time that an action is required
according to the
Blue Status Light,
then by default the yellow status light will
dominate.
Table 8.
Fuses Parameters
Parameter
Value
Rated Current
6.3 A
Rated Voltage
250VAC
Fuse Dimensions
5 x 20 mm
Table 7.
Status LED Lights (Sheet 2 of 2)
Symbols
Description