3
Setup Menu Item
Setup Menu Item Modules
130
TriPlus RSH User Guide
Thermo Scientific
Max Temperature
Sets the upper temperature limit.
The Maximum Temperature which can be set is 350 °C, but
this parameter is often defined by the specific needs for the
application.
Min Temperature
Sets the lower temperature limit.
Standby Temperature
The Standby Temperature can be adjusted to the
application’s requirements. Setting the Standby Temperature
using this function does not yet activate the newly-selected
temperature. To activate it, go to the pull-up menu
Options
and select
Activate Standby Temperature
.
The Standby Temperature is activated when the TriPlus
RSH is powered up. In routine use, the Standby
Temperature is activated after a job has been terminated.
The Standby Temperature is typically lower than the
application temperature.
Default Needle Penetration
Depth
Defines the penetration depth of the SPME Needle in the
module. The value reflects the total length including the
length of the fiber.
Max Needle Penetration
Depth
Upper limit of penetration depth of the SPME needle in the
module. The value reflects the total length including the
length of the fiber.
Temp. Gain
The coefficient used, if a temperature calibration has been
performed. Possible values are between 0 and 2. Read-only.
Temp.Offset
The Offset temperature used, if a temperature calibration
has been performed. Possible values are between -10 °C and
+10 °C. Read-only.
Table 60.
SPME Conditioning Station Parameters (Sheet 2 of 2)
Item
Description