Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
105
Chromatograph System Options Pull-up Menu
Selecting the pull-up menu
Options
opens up the following tasks. See
.
Table 37.
Chromatography System Parameters
Item
Description
Cryo Trap
Sets start signal for the Cryo Trap, Output signal required.
Fake Injection Delay
Sets the delay time until the start of a fake injection.
A fake injection performs a start of the chromatographic
system, without injection of a sample.
Injected
Sets start signal for the chromatographic system, for example
GC System1. Output Signal is required.
Ready
A ready signal from another device is expected as an Input
Signal.
Note
The synchronization between the TriPlus RSH and any chromatographic system has
to be checked for each particular system configuration. The setting of the synchronization
signal has to be matched with the pin allocation of the synchronization cable.
The original cable are labeled. Check the labels first before connecting. This is especially
important with the Y-cable configurations where two GC units are connected to a single
TriPlus RSH.
Table 38.
Chromatography System Pull-up Menu Options
Item
Description
New Chromatograph System
Creates a new Chromatograph System.
Copy GC
Copies the Chromatograph System, GC1 as the example.
Delete GC
Deletes the Chromatograph System; GC1 as the example.