5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
201
Upon starting the wizard, the TriPlus RSH automatically turns off the current to enable a
manual movement of the TriPlus RSH Head.
5. In case the calibration step is activated with an installed tool, the system dialog will ask
you to move the head first to an empty slot, and
Eject
the tool first.
When Step 4 has been successfully completed, press
Next
to continue to step 5 to
calibrate the coupling.
6. In Step 6, the Sensor of the
Plunger motor
is calibrated. When finished, press
Next
to
continue.
Note
Please note that the TriPlus RSH Head has to be positioned such that the
receiving holes of the
Z-Slider
are above the
Coupling Pivots
of the tool but they
must not be engaged.
Note
When the tool ejection has been completed, the tool is not installed in the head
of the TriPlus RSH. Continuing with
Calibrate Motors
, in the Installation Wizard
will detect that the tool is not mounted, and will adapt to this situation accordingly.