The tray type parameters are described in the following
Z Tolerance
A tolerance window to give a plus/minus range (expressed in
mm) where the Head must expect an object.
If the value of this item is
> 0
, indicates a relative detection
mode, and the sensor from the Head needle guide is active.
If the value of this item is set to
0
, the Needle Guide Motor
detection current is turned off, and the syringe slider moves
to an absolute value. The Z-axis position as specified by the
object.
The Z Tolerance is calculated in combination with a Rack
from the Vial top. The Rack plate top surface is not used as
reference.
The Z Tolerance is calculated in combination with a Well
Plate (MTP) from the plate top surface. Unit: mm.
DefNdlPenetrDepth
(Parameter used for MTP
only)
Defines the default value for the syringe needle penetration
depth in the well of the Micro Titer Plate (MTP). Unit: mm.
Depenetration Speed
(Parameter used for MTP
only)
Defines the speed of the syringe needle de-penetration from
the well of the Micro Titer Plate (MTP). Unit: mm/s.
Sealing Foil
(Parameter used for MTP
only)
Setting the mark to
Has Foil
indicates that the plate is sealed
with a foil. The
Foil Piercing Tool
will be applied.
MaxVolume
(Parameter used for MTP
only)
Defines the maximum volume of the well of a Micro Titer
Plate (MTP). Unit: mL.
Penetration Speed
(Parameter used for MTP
only)
Defines the speed of the syringe needle penetration from the
well of the Micro Titer Plate (MTP). Unit: mm/s.
Bottom Sense supported
(Parameter used for MTP
only)
Setting the mark to
Bottom Sense supported
enables the
functionality of needle bottom sensing.