Setup Menu Item
Setup Menu Item Modules
Thermo Scientific
TriPlus RSH User Guide
111
Figure 80.
OC Actuator - Selection and Parameters
Press
Enter
to edit a parameter. See
Table 43.
OC Actuator Parameters (Sheet 1 of 2)
Item
Description
Def Ndl Penetr Depth
Allows you to define the
Needle Penetration Depth
in the
OC injector.
Opened
Changes the status
Opened
or
Closed
.
This parameter is a check if the
gas lock
is opened or closed.
If marked as true (check box marked) the injector is defined
as
open
. This can be verified by inserting a syringe manually
and if a resistance is noted, the gas lock is closed.
If a mismatch is observed, check the settings of the
parameters
Homing Direction
and
Homing Position
.
Actual Position
Reads out the position of the OC Actuator. This information
can be used for troubleshooting purposes.
Homing Direction
Changes the direction of homing.
Homing Position
Changes the position of homing.
Max Ndle Penetr Depth
Defines the maximum allowed syringe needle penetration in
millimeters.