TriPlus RSH User Guide
199
•
Corporate Network with DHCP Protocol
— Ask the network administrator to verify
and integrate the setup within the corporate network.
Select the DHCP Protocol; the network will assign an IP Address and a Subnet Mask to
the TriPlus RSH.
•
Corporate Network with Fixed IP Address
— Ask the network administrator to provide
the IP Address, Subnet Mask, and Default Gateway to configure the TriPlus RSH within
the corporate network.
Calibrate Motors
The calibration of the motors is basically a check to determine whether the mechanical stop
points of the various axes can be found reliably. A clearance distance (offset) from the
mechanical stop point is assigned and stored as the electronic stop point. This calibration of
the axes is mandatory and ensures reliable performance.
This step has been repeated when a motor has failed, has been replaced or when any other
major error has occurred. This calibration task brings the TriPlus RSH back to reliable
operation.
To calibrate motors
1. Select the wizard item
Calibrate Motors
and follow the dialog. In the following example,
the Calibration Wizard guides you from the steps 1 to 6. See
Figure 141.
Wizard Item Calibrate Motors and Starting Calibration Wizard Step 1 of 6
Press
Calibrate
to start the first step
Calibrate Axes
. All axes from the TriPlus RSH are
calibrated. Wait until the yellow light stops flashing and
Next
becomes available.
Note
Please do not confuse the terms and tasks for
Calibration Motors
as explained in
this section with those for
Calibration Tools
as described in the section
When starting the wizard
Calibrate Motors
at least one tool must be parked
in a slot of the Tool Station.