5
Service Menu Item
Installation Item
210
TriPlus RSH User Guide
Thermo Scientific
Figure 152.
Set Reference Point and Finish Wizard
Calibrate Tools
Lunette
Note
The concept of
Referencing
and
Calibration
is explained in the chapter
TriPlus
RSH Defining Object Positions
in the
TriPlus RSH Hardware Manual
.
The tool which is most frequently used should be used for Calibration.
This tool can be considered as the
ideal
tool. All other tools will be referenced with respect
to the
Ideal
tool.
This step is not necessary when a single tool station, the TR Station, is used or if only one
tool is in use within the entire system.
CAUTION
It is advisable to hold the Y-Axis with one hand and with the other hand hold
the tool body to move the TriPlus RSH Injection Unit in a controlled way to the teaching
position.
Do NOT hold the Needle Guide to move the TriPlus RSH Injection Unit. This part is
mechanically flexible and could lead to misalignment.