Item
Description
Max Penetration Depths
The maximum allowed Needle Penetration can be defined as
a safety measure so that you cannot destroy the needle tip by
going too far down.
The maximum needle penetration must not exceed the
length of the syringe needle, considering the loss of
approximately 12 mm due to the needle guide, and other
factors. For a typical needle length of 57 mm or 85 mm for
the long needle, the maximum penetration depth must not
exceed 45 mm or 73 mm.
Def Nedl Penetr Depth
The default value of the syringe needle penetration depth in
the reservoir vial is determined by this parameter.