1
Getting Familiar with Your TriPlus RSH
Barcode Reader
Thermo Scientific
TriPlus RSH User Guide
67
Figure 48.
Examples of Barcode Label Positioning and Label Size
Reading the Barcode from a Vial or Well Plate
The Barcode Reader reads a barcode label from an individual vial, but it is not able to scan
and treat multiple units of information, as would typically be received from a multiple well
plate.
Multidimensional Barcodes
The scan engine used for the Barcode Reader Module is designed as a one-dimensional (1D)
scanner. Two-dimensional barcodes (2D) cannot be decoded.
Barcode Label Size
The minimum length of the barcode is 18 mm if used with a 20 mL Vial. Please note that this
refers to the actual barcode length and not the label itself.
Barcode Reader Specifications
The Barcode Reader Specifications are:
•
Vial Sizes
— 2 mL, 10 mL, and 20 mL
•
Electrical Connectors
— Two BUS connectors for daisy chaining.
The Barcode Reader dimensions, mass, operating and environmental requirements are
detailed in
and
.