3
Setup Menu Item
Setup Menu Item Modules
168
TriPlus RSH User Guide
Thermo Scientific
Wash Stations
Three different wash stations are available with the TriPlus RSH:
Standard Wash Station
(passive module),
Large Solvent Wash Station
(passive module), and
Fast Wash Station
(active module).
Selecting the list item
Wash Stations
opens up the following parameters:
Figure 126.
Menu Item Wash Stations Menu and List Items Wash Modules
Press
Enter
to edit a parameter.
Table 79.
Options Vial Type Pull-up Menu
Item
Description
New Vial
Creates of a new Vial.
Copy Vial
Copies the present Vial.
Delete Vial
Deletes the present Vial.
Reset to Default
Resets the selected Vial parameter to default value.
Check Teaching
Checks the teaching position of the module.
This task serves as a quick check to determine if the X-, Y-,
and Z-positions are defined correctly.
In case a teaching point is off, return to the Menu
Setup
for
readjustment of the teaching position. Please note that this
action can be done under the
Extended User Level
but not
under
User Level
.
Teach Module
The positions of the X-, Y-, and Z-axes can be defined again,
(teaching).