T a b le 5 (c o n t)
Characteristics
Performance Requirements
Supplemental Information
X-Y OPERATION (cont)
Bandwidth
X-Axis
Dc to at least 2 MHz.
Y-Axis
Dc to at least 100 MHz.
Input Characteristics
Resistance
1 M fi ±2%.a
Capacitance
20 pF ±10%.a
Phase Difference Between
X- and Y-Axis Am plifiers
< 3° from dc to 200 kHz.
Accuracy
X-Axis
0°C to +40° C
±5% o f indicated deflection.
— 15° C to +55° C
±8% o f indicated deflection.3
Perform ance Requirement not checked in Service Manual.
@