Substituting these values in the vertical conversion factor
formula:
Vertical
Conversion
Factor
30 V
4 div x 5 V/div
1.5
Continuing, for the unknown signal the VO LTS/DIV
switch setting is
1
and the peak-to-peak amplitude spans
five vertical divisions. The arbitrary deflection factor is then
determined by substituting values in the formula:
reference signal (e.g., on assembly line test) may be easily
and accurately measured w ith the 2337. To accomplish
this, a reference signal o f known time duration is first set
to an exact number o f horizontal divisions by adjusting the
A SEC/DIV switch and the TIME (PULL) VAR control.
Unknown signals can then be compared w ith the reference
signal w ithout disturbing the setting o f the TIME (PULL)
VAR control. The procedure is as follows:
1.
Set the time duration o f the reference signal to an
exact number o f horizontal divisions by adjusting the A
SEC/DIV switch and TIME (PULL) VAR control.
Arbitrary
Deflection = 1.5 x 1 V/div = 1.5 V/div
Factor
2.
Establish a horizontal conversion factor, using the
follow ing formula (reference signal time duration must be
known):
The amplitude o f the unknown signal can then be deter
mined by substituting values in the unknown signal ampli
tude formula:
Am plitude = 1.5 V /div x 5 div = 7.5 V
Horizontal
Conversion
Factor
reference signal time
duration (seconds)
horizontal
A SEC/DIV
distance
x
switch
(divisions)
setting
Time Comparison
In a similar manner to "A m plitude Comparison",
repeated time comparisons between unknown signals and a
3.
For the unknown signal, adjust the A SEC/DIV
switch to a setting that provides sufficient horizontal deflec
tion to make an accurate measurement. Do not readjust the
TIME (PULL) VAR control.
@
2237 Operators
55
Содержание 2337
Страница 9: ... ...
Страница 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Страница 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Страница 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Страница 72: ... Figure 22 Delayed sweep magnification ...
Страница 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Страница 116: ...NOTES ...