2.
Set the B TRIGGER SOURCE switch to RUNS
AFTER DLV.
3.
Apply the signal to either vertical-channel input
connector and set the VE R TIC A L MODE switch to display
the channel used.
4.
Set the appropriate VO LTS/DIV switch to produce a
display o f approximately 5 divisions in amplitude and
center the display.
5.
Set the A SEC/DIV switch to a sweep speed which
displays at least one complete waveform cycle.
6
.
Select the A INTEN HORIZ MODE and adjust the
B DELAY TIME POSITION control to position the
intensified zone to the portion o f the display to be mag
nified (see Figure 22A).
7.
Set the B SEC/DIV switch to a setting which inten
sifies the fu ll portion o f the A trace to be magnified. The
intensified zone w ill be displayed as the B trace. The start
o f the intensified zone remains as previously positioned
in step
6
.
58