T a b le 5 (c o n t)
Characteristics
Performance Requirements
Supplemental Information
V E R TIC A L DEFLECTION SYSTEM (cont)
Step Response (cont)
Rise Time (5 mV per division to
5 V per division)
— 15° C to +40° C
3.5 ns or less.
Rise time is calculated from the
formula:
, .
350
Rise Time (ns
=
—
. ,
BW (MHz)
+40° C to +55° C
4.15 ns or less.3
Aberrations
Positive-Going Step
(Excluding ADD Mode)
5 mV per division to 0.2 V
per division
+3%, —3%, 3% p-p or less.
Negative-Going Step
Add 2% to all positive-going step
specifications; checked at 5 mV
per division.
Perform ance Requirement not checked in Service Manual.
@