27.
Position Channel 1 trace tw o divisions above the
center graticule line.
28.
Select CH 2 VER TICAL MODE and position the
Channel 2 trace tw o divisions below the center graticule
line.
29.
Select A L T VER TICAL MODE and rotate the A
SEC/DIV switch throughout its range. The display w ill
alternate between channels at all sweep speeds. This mode
is most useful fo r sweep speeds from 0.05
ns
to 0.2 ms per
division.
30.
Select CHOP VER TICAL MODE and rotate the A
SEC/DIV switch throughout its range. A dual-trace display
w ill be presented at all sweep speeds, but unlike the A L T
mode, both Channel 1 and Channel 2 signals are displayed
fo r each sweep speed on a time-shared basis. This mode is
most useful for sweep speeds from 0.5 ms to 0.5 s per
division.
31.
Set the A TRIGGER SOURCE switch to VERT
MODE.
32.
Observe the switching between Channel 1 and
Channel 2 at the higher sweep speeds (indicated by the
segmented trace).
33.
Select AUTO V E R TIC A L MODE by simultaneously
pressing in both A L T and CHOP push buttons.
34. Rotate the A SEC/DIV switch throughout its range.
Note that a dual-trace display is present at all sweep speeds.
35. Select CH 1 V E R TIC A L MODE and set Channel 1
AC-GND-DC switch to DC. Recenter the display on the
screen.
Using the Horizontal Section
1.
Return the A SEC/DIV switch to 0.5 ms and note
the display fo r future comparison in step 3.
2. Set the A SEC/DIV switch to 5 ms and press in the
X10 MAG push button.
3. Observe that the display is similar to that obtained
in step 1.
30
2337 Operators
@
Содержание 2337
Страница 9: ... ...
Страница 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Страница 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Страница 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Страница 72: ... Figure 22 Delayed sweep magnification ...
Страница 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Страница 116: ...NOTES ...
Страница 117: ...NOTES ...