BASIC APPLICATIONS
A fter becoming familiar w ith all the capabilities o f the
2337 Oscilloscope, the operator can then adopt a con
venient method for making a particular measurement. The
following information describes the recommended pro
cedures and techniques for making basic measurements
w ith your instrument. When a procedure first calls for
presetting instrument controls and obtaining a Normal
Sweep Display, refer to the "Instrum ent Fam iliarization"
section and perform steps 1 through 4 under "Normal
Sweep Display."
NONDELAYED MEASUREMENTS
AC Peak-to-Peak Voltage
To perform a peak-to-peak voltage measurement, use the
following procedure:
N O T E
This procedure may also be used to make voltage
measurements between any two points on the
waveform.
1.
Preset instrument controls and obtain a Normal
Sweep Display.
2.
A pply the ac signal to either vertical-channel input
connector and set the VE R TIC A L MODE switch to display
the channel used.
3.
Set the appropriate VO LTS/D IV switch to display
about five divisions o f the waveform, ensuring that the
VO LTS/D IV VAR control is in the calibrated detent.
4.
Adjust the A TRIGGER LEVEL control to obtain
a stable display.
5. Set the A SEC/DIV switch to a position that displays
several cycles o f the waveform.
6. Vertically position the display so that the negative
peak o f the waveform coincides w ith one o f the horizontal
graticule lines (see Figure 13, Point A).
40
2337 Operators
@
Содержание 2337
Страница 9: ... ...
Страница 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Страница 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Страница 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Страница 72: ... Figure 22 Delayed sweep magnification ...
Страница 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Страница 116: ...NOTES ...
Страница 117: ...NOTES ...