) TRIG HOLDOFF (PUSH) VAR Control-Provides
continuous control o f h o ld o ff time between sweeps.
This control improves the ability to trigger on aperi
odic signals (such as complex digital waveforms)
and increases the minimum h o ld o ff time to at least
2.5 times at any sweep speed.
B TRIGGER
Refer to Figure 7 fo r location o f items 35 through 41.
(35) LEVEL C ontrol—Selects the amplitude point on the
trigger signal at which the sweep is triggered. This
control is usually adjusted fo r the desired display
after Trigger SLOPE and SOURCE switch settings
have been selected.
(36) SOURCE Switch—Determines the mode o f operation
fo r the B Sweep and the signal source for the B
Trigger.
A TIM E —Provides tw o intensified zones on the
c rt trace fo r differential time measurements. The
time difference between the tw o intensified zones
is determined by the B DELAY TIM E POSITION
and the A TIM E POSITION controls. Time differ-
@