Figure 20. Phase difference.
Figure 21. High-resolution phase difference.
Figure 21 shows the same signals illustrated in Figure 20,
but magnifying the displays results in a horizontal
difference o f
6
divisions between the tw o signals.
Substituting the given values in the phase difference
formula:
Phase Difference =
6
div x 4 .5 °/div = 27°
Amplitude Comparison
In some applications it may be necessary to establish
a set o f deflection factors other than those indicated by
the VO LTS/DIV switch settings. This is useful fo r com
paring unknown signals to a reference signal of known
amplitude. To accomplish this, a reference signal o f known
amplitude is first set to an exact number o f vertical divisions
by adjusting the VO LTS/DIV switch and VAR control.