o f the waveform touches the 100% graticule line (see
Figure 18).
Substituting the given values in the formula:
6. Set the A SEC/DIV switch fo r a single-waveform
display, w ith the rise time spread horizontally as much as
possible.
Rise Time
5 div x 1 yus/div
1
= 5 /is
7. Horizontally position the display so the 10% point
on the waveform intersects the second vertical graticule
line (see Figure 18, Point A).
8.
Measure the horizontal distance between the 10% and
90% points and calculate the time duration using the
following formula:
horizontal
SEC/DIV
distance
x
switch
_.
(divisions)
setting
Rise Time = -------------—— :— -------------
magnification factor
EXAMPLE: The horizontal distance between the 10%
and 90% points is 5 divisions (see Figure 18), and the A
SEC/DIV switch is set to 1 /ts/division. A magnification
factor o f 1 is used.
Figure 18. Rise time.
@
2337 Operators
49