DELAYED-SWEEP MAGNIFICATION
The delayed-sweep feature o f the 2337 can be used to
provide higher apparent magnification than is provided by
the X I 0 MAG function. Apparent magnification occurs as
a result o f displaying a selected portion o f the A trace at
a faster sweep speed (B Sweep speed). The A SEC/DIV
switch setting determines how often the B trace w ill be
displayed. Since the B Sweep can occur only once fo r each
A Sweep, the A Sweep time duration sets the amount o f
time elapse between succeeding B Sweeps.
The intensified zone is an indication of both the location
and length of the B Sweep interval w ithin the A Sweep
interval. Positioning o f the intensified zone (i.e., setting
the amount o f time between start o f the A Sweep and start
o f the B Sweep) is accomplished w ith the B DELAY TIME
POSITION control. With either the A INTEN or the B
HORIZ MODE selected and w ith the B TRIGGER
SOURCE switch set to RUNS AFTER D LY, the B D ELAY
TIME POSITION control provides continuously variable
positioning of the start o f the B Sweep. The range o f this
control is sufficient to place the B Sweep interval at any
location w ithin the A Sweep interval. When the A INTEN
HORIZ MODE is selected, the B SEC/DIV switch setting
determines the B Sweep speed and concurrently sets the
length o f the intensified zone on the A trace.
Using delayed-sweep magnification may produce a
display w ith some slight horizontal movement (pulse jitte r).
Pulse jitte r includes not only the inherent uncertainty of
triggering the delayed sweep at exactly the same trigger
point each time, but also jitte r that may be present in
the input signal. If pulse jitte r needs to be measured, use
the "Pulse Jitter Time Measurement" procedure which
follows the discussion o f "Magnified Sweep Runs A fter
Delay."
When looking at a signal having pulse jitte r, it is often
useful to be able to hold that signal stationary on the
screen. Operating the B Sweep in a triggered mode w ill
produce a more stable display, since the delayed display is
triggered at the same level each time. This capability is
provided by the 2337 and is explained in the "Triggered
Magnified Sweep" discussion.
Magnified Sweep Runs After Delay
The following procedure explains how to operate the
B Sweep in a nontriggered mode and to determine the
resulting apparent magnification factor.
1.
Preset instrument controls and obtain a Normal
Sweep Display.
@
2237 Operators
57
Содержание 2337
Страница 9: ... ...
Страница 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Страница 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Страница 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Страница 72: ... Figure 22 Delayed sweep magnification ...
Страница 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Страница 116: ...NOTES ...
Страница 117: ...NOTES ...