T a b le 5 (c o n t)
Characteristics
Performance Requirements
Supplemental Information
H O RIZO NTAL DEFLECTION SYSTEM (cont)
Linearity
±5%.
Over any 2-division portion of the full
10 divisions, displayed at all sweep
speeds. Exclude the first and last dis
played divisions of the 5- and 10-ns
per division sweep speeds w ith X10
MAG on.
Variable Range (VAR)
Continuously variable between calibrated
settings of the SEC/DIV switches.
Extends maximum A Sweep speed to
at least 1.25 s per division.
A Sweep Length
10.5 to 11.5 divisions.
Checked at 1 ms per division.
A Trigger H oldoff (VAR)
A t least 2.5 times the minimum h oldoff
at any sweep speed.3
Magnifier Registration
±0.2 division from graticule center (X10
MAG on to X10 MAG off).
POSITION Control Range
Start o f sweep must position to right of
graticule center. End o f sweep must
position to left o f graticule center.
Checked at 1 ms per division.
Perform ance Requirement not checked in Service Manual.
@