. Select the B HORIZ MODE to magnify the inten
sified portion o f the A Sweep (see Figure 22B).
2. Referring to Figure 23, measure the difference
between Point A and Point B in divisions and calculate
the pulse jitte r time using the following formula:
9. The apparent sweep magnification can be calculated
from the following formula:
Pulse
Jitter
Time
horizontal
difference
x
(divisions)
B SEC/DIV
switch
setting
Apparent
Delayed Sweep
Magnification
A SEC/DIV switch setting
B SEC/DIV switch setting
EXAMPLE: Determine the apparent magnification o f a
display w ith an A SEC/DIV switch setting o f 0.1 ms and
a B SEC/DIV switch setting o f 1
ms
-
Substituting the given values:
Apparent
1 x 10
’ 4
s
=
----------------------------- —
=
102
=
100
Magnification
1 x 1 0 s
Pulse Jitter Time Measurement
1. Perform steps 1 through
8
o f the preceding "Mag
nified Sweep Runs A fte r Delay" procedure.
M E A S U R E
T IM E F R O M
® TO ®
4119-24
Figure 23. Pulse jitter.
@