Table 6 (cont)
Characteristics
Description
Vibration (Operating)
15 minutes along each o f 3 major axes at a total displacement o f 0.025 inch
p-p (4 g at 55 Hz), w ith frequency varied from 10 Hz to 55 Hz to 10 Hz in
1-minute sweeps. Hold 10 minutes at each major resonance, or if none exists,
hold 10 minutes at 55 Hz (procedure differs from MIL-T-28800B).
Shock (Operating and Nonoperating)
50 g, half-sine, 11-ms duration, 3 shocks per axis in each direction, fo r a total
o f 18 shocks.
EMI
Will meet MIL-STD-461A requirements using procedures outlined in MIL-STD-
462, except where 10 V/m is used in place of 1 V/m.
Transportation
Meets the limits o f National Safe Transit Association test procedure 1A-B
w ith a 36-inch drop.
Humidity
Operating
Oscilloscope
DMM
+ 55°C, 90% relative humidity, for at least 72 hours.
as + 35°C, 90% relative humidity, for at least 24 hours.
< + 55°C, 70% relative humidity, for at least 24 hours.
Nonoperating
(Storage) (both)
+ 60°C, 90% relative humidity, for at least 72 hours.
9 6
2337 Operators
REV JAN 1982