DM M DISPLAYS AND
MEASUREMENTS
Before connecting the DMM test leads to a unit-under-
test,
select
the desired measurement
RANGE and
FUNCTION. When the value o f the quantity to be
measured is unknown, select the highest range first.
Decrease the range setting until the display reads between
10% and 100% o f the full-range value.
Resistance
The D M M m ay be damaged i f i t is operating in the
resistance m ode (SI fu n c tio n selected) and m ore than
5 0 0 V is applied between the + (red) and - (black)
inputs.
1.
Rotate the RANGE/FUNCTION switch to any
SI
range.
The m axim um safe in p u t voltage is 5 0 0 V (dc + peak
ac) between the + (positive) a n d — (negative) in p u ts
o r between the + an d — in p u ts to ground.
If an overrange condition exists, a " 1 " w ill be displayed
on the left side o f the LCD readout, followed by three
blank digits and a decimal point. Should this occur, select
the next higher RANGE value.
The 500 V DC and 350 V RMS AC RANGE w ill not
indicate an overrange condition when the input signal
exceeds the maximum allowable input voltage.
2.
Connect the + (red) and — (black) leads across
the unknown resistance.
3.
Select the best
SI
range to obtain maximum resolu
tion w ith o u t overranging; read the resistance value on the
LCD readout.
RMS AC
I f the LC D readout exceeds 3 5 0 V, im m ediately
disconnect the test leads fro m the unit-under-test
to prevent possible in stru m e n t damage.
2237 Operators
67
Содержание 2337
Страница 9: ... ...
Страница 20: ...4119 06 Figure 5 Horizontal controls and indicator ...
Страница 24: ...13 ire 6 A TRIGGER controls connector and indicator ...
Страница 29: ...Figure 7 DMM and B TRIGGER controls connectors and LCD readout 16 2337 Operators ...
Страница 72: ... Figure 22 Delayed sweep magnification ...
Страница 81: ...Figure 27 Time difference between two time related pulses 66 2237 Operators ...
Страница 116: ...NOTES ...
Страница 117: ...NOTES ...