Table 1
Line Voltage and Fuse Selection
LINE FUSE
Line Voltage
Selector Switch
Voltage
Position
Range
Fuse Data
115 V Nominal
100 to 132 V
1.0 A, 250 V, Fast-blow
230 V Nominal
200 to 250 V
0.5 A, 250 V, Fast-blow
Table 2
Option 03 Line Voltage and Fuse Selection
Line Voltage
Selector Switch
Voltage
Position
Range
Fuse Data
100 V Nominal
90 to 115 V
1.0 A, 250 V, Fast-blow
200 V Nominal
180 to 230 V
0.5 A, 250 V, Fast-blow
To verify that the instrument power-input fuse is o f
proper value fo r the nominal ac source voltage, perform
the following procedure:
1. Press in the fuse holder cap and release it w ith a
slight counterclockwise rotation.
2. Pull the cap (with the attached fuse inside) out o f
the fuse holder.
3. V erify proper fuse value (see Tables 1 and 2).
4
2337 Operators
@
Содержание 2337