Table 5 (cont)
Characteristics
Performance Requirements
Supplemental Inform ation
H O RIZO NTAL DEFLECTION SYSTEM (cont)
Differential Time Measurement
Accuracy
+15°C to +35° C
±1% o f reading ±1 count.
Exclude delayed operation when
knobs are locked at any sweep speed
or when the A SEC/DIV switch is at
either 0.1 /us per division or 0.05 /us
per division. Exclude the first 0.25
division on all A Sweep speeds.
—15°C to +55°C
±2.5% o f reading ±1 count.8
Delay Time Jitter
±0.005% o f 10 times the A SEC/DIV
switch setting (less than one part in
20,000) over the fu ll delay time range.
X-Y OPERATION
Deflection Factor Range
5 mV per division to 5 V per division
in a 1, 2, 5 sequence.
No X-axis variable.
Performance Requirement not checked in Service Manual.
8 2