dc level than that previously applied, especially if the dc-
level difference is more than 10 times the VO LTS/DIV
switch setting:
4. Connect the probe tip to the signal source.
1. Set the AC-GND-DC switch to GND before connect
ing the probe tip to a signal source.
5. Wait several seconds fo r the input coupling capacitor
to charge.
N O T E
The outer shells o f the A EXT, CH 1 OR X, and
CH 2 OR Y connectors are attached to the 2337
chassis ground.
6. Set the AC-GND-DC switch to AC. The display w ill
remain on the screen, and the ac component o f the signal
can be measured in the normal manner.
2. Touch the probe tip to the oscilloscope chassis
ground.
INSTRUMENT COOLING
3. Wait several seconds fo r the input coupling capacitor
to discharge.
To maintain adequate instrument cooling, the ventila
tion holes on both sides o f the equipment cabinet must
remain free o f obstructions.